Skip to content

Our fluorescence spatial resolution targets align with the FDA Draft Guidance and AAPM TG311 Report on standardizing benchmarks for fluorescence surgical imaging system performance. 

The spatial resolution target provides positive fluorescence contrast to benchmark fluorescence imaging system performance with a standardized 1951 USAF resolution target mask atop fluorescent ICG-01 backing material. This target is helpful for:

  • characterize fluorescence lateral spatial resolution
  • empirically characterizing fluorescence imaging depth of field
  • compare predicate device imaging performance for equivalence justification (e.g., 510(k) submission)

We provide a getting started guide, directions for use, and practical tutorials on using fluorescence spatial resolution targets for your application. 

Our open-source QUEL-QAL python library helps you generate standardized performance reports from raw image data, enabling consistent benchmarking across development cycles and facilitating objective system-to-system comparisons. Video tutorials for using QUEL-QAL can be found in this playlist on our YouTube page.

To help maintain the performance and longevity of the fluorescent reference targets, we recommend minimizing prolonged light exposure. For absolute radiometric benchmarking of system performance, we recommend using the RET target.