ICG Concentration Sensitivity Target
9-well fluorophore concentration target for fluorescence sensitivity performance characterization in tissue. Includes electronic Certificate of Verification (.PDF).
Fluorophore: ICG-01 (Peak Ex/Em: 811/822 nm)
Suitable for applications using 820 nm channel fluorophores (e.g., ICG, indocyanine green)
Fluorescent Material: ICG-01 = 1.0 - 1000 nM
Optical Properties measured at 𝛌 = 800 nm
- Absorption (µₐ) = 0.03 mm⁻¹
- Reduced scattering (µ′ₛ) = 0.4 mm⁻¹
Standard Size Dimensions:
- 50 x 50 x 20 mm
- ⦰ 10 mm feature size
Mini Size Dimensions:
- 25 x 25 x 20 mm
- ⦰ 5 mm feature size
Compatible with our open-source QUEL-QAL image analysis library.
Fluorophore: ICG-01 (Peak Ex/Em: 811/822 nm)
Suitable for applications using 820nm channel fluorophores (e.g., ICG, indocyanine green)
Fluorescent Features:
- [1, 3, 10, 30, 60, 100, 300, 1000] nM of ICG-01
- 1 negative control well (0 nM)
Base optical properties:
- Absorption (µₐ) = 0.03 mm⁻¹ at 800nm
- Reduced scattering (µ′ₛ) = 0.4 mm⁻¹ at 800nm
Dimensions:
- (Standard): 50x50x20mm | ⦰10mm feature size
- (Mini): 25x25x20mm | ⦰5mm feature size
Our fluorescence concentration sensitivity targets align with international consensus established through AAPM Task Group Report 311 and the draft FDA guidance on developing drugs for optical imaging.
The ICG Concentration Sensitivity Target simplifies characterizing your system's critical fluorescence imaging performance by assessing:
- linear dynamic range
- noise floor
- detection limits
- saturation threshold
The concentration target is useful for establishing automated imaging parameters, optimizing image processing pipelines, and verifying system performance over a ~1000-fold concentration range.
We provide a getting started guide, directions for use, and practical tutorials on how to get the most out of your depth sensitivity target for your application.
Our open-source QUEL-QAL python library helps you generate standardized performance reports from raw image data, enabling consistent benchmarking across development cycles and facilitating objective system-to-system comparisons. Video tutorials for using QUEL-QAL can be found in this playlist on our YouTube page.
To help maintain the performance and longevity of the fluorescent reference targets, we recommend minimizing prolonged light exposure. For absolute radiometric benchmarking of system performance, we recommend using the RET target.