Quantitative Q800 Reference Set
Absolute fluorescence performance characterization against an included calibrated radiometric reference standard.
Reference set includes:
- Q800 Concentration Sensitivity Target
- Q800 Depth Sensitivity Target
- Q800 Resolution Target
- 800 nm Radiometric Target with Power Supply
- Electronic Certificate of Verification (.PDF) and Calibration File (.XLS)
Sizes:
- Standard: 50 x 50 x 20 mm
- Mini: 25 x 25 x 20 mm
Fluorophore: Q800-01 (Peak Ex/Em: 775/787 nm)
Suitable for systems optimized for 800 nm channel fluorophores (pafolacianine, OTL38, IRDye® 800CW)
Compatible with our open-source QUEL-QAL image analysis library.
Fluorophore: Q800-01 (Peak Ex/Em: 775/787 nm)
Suitable for systems optimized for 800 nm channel fluorophores (pafolacianine, OTL38, IRDye® 800CW)
Q800 Concentration Sensitivity Target
Optical Properties:
Optical Properties measured at 𝛌 = 800 nm
- Absorption (µₐ) = 0.04 mm⁻¹
- Reduced scattering (µ′ₛ) = 1.2 mm⁻¹
Fluorescent Features:
- [1, 3, 10, 30, 60, 100, 300, 600] nM of Q800-01
- 1 negative control well (0 nM)
Base optical properties:
- Absorption (µₐ) = 0.04 mm⁻¹ at 800 nm
- Reduced scattering (µ′ₛ) = 1.2 mm⁻¹ at 800 nm
Dimensions:
- (Standard): 50x50x20mm | ⦰10mm feature size
- (Mini): 25x25x20mm | ⦰5mm feature size
Q800 Depth Sensitivity Target
Optical Properties:
Top: 0.5-6mm tissue-equivalent material non-fluorescent material
- Absorption (µₐ) = 0.01 mm⁻¹ at 800 nm
- Reduced scattering (µ′ₛ) = 1.5 mm⁻¹ at 800 nm
Bottom: 8 wells 600 nM Q800-01
- Absorption (µₐ) = 0.04 mm⁻¹ at 800 nm
- Reduced scattering (µ′ₛ) = 1.2 mm⁻¹ at 800 nm
Negative control has 6mm of top material
Dimensions:
- (Standard): 50x50x20 mm | ⦰10 mm feature size
- (Mini): 25x25x20 mm | ⦰5 mm feature size
Q800 Resolution Target
Target Mask: Negative 1951 USAF Resolution Target
- Chrome-on-glass, reflective contrast mask
- AR-coated for NIR illumination
- Group 0 through 7 contrast features
- Minimum feature size: 228 lpmm || 2.19 µm line spacing
Fluorophore: 600 nM Q800-01 (Peak Ex/Em: 775/787 nm)
Optical Properties
- Absorption (µₐ) = 0.04 mm⁻¹ at 800 nm
- Reduced scattering (µ′ₛ) = 1.2 mm⁻¹ at 800 nm
Dimensions:
- 25x25x20 mm
800 nm Radiometric Emitter Target (RET)
Optical Output:
- Center Wavelength = 800 nm
- Adjustable radiometric output from ~0.05 - 300 µW/(cm^2 str)
- Mimics NIR fluorescence spectral emission properties
- Calibration file provided for each unit.
Dimensions:
- 25x25x20mm
Guides & Resources:
Concentration Target Getting Started Guide
Depth Target Getting Started Guide
Resolution Target Getting Started Guide
Radiometric Emitter Target Getting Started Guide
Radiometric Emitter Application Note - Photobleaching Management
Reference Target Directions for Use
FAQs
Technical Drawings
Standard Concentration Sensitivity Target Technical Drawing
Standard Depth Sensitivity Target Technical Drawing
Resolution Target Technical Drawing
RET Technical Drawing
Our Q800 quantitative reference set adds a traceable, calibrated radiometric emitter to our base fluorescence reference set. The emitter provides a stable reference optical signal that is not impacted by illumination conditions, enabling traceable optical characterization across multiple imaging systems for standardized comparison.
A stable, calibrated reference signal lets you turn your quantized image intensities into standardized radiometric units (e.g., mW/cm^2 str). The flexibility of a stable reference signal helps dial in imaging parameters and standardize traceable performance characterization in absolute radiometric units.
Our base fluorescence reference target kit aligns with the FDA Draft Guidance and AAPM TG311 Report on standardizing benchmarks for fluorescence surgical imaging system performance. This kit enables teams to characterize critical system performance metrics including:
- fluorescence sensitivity range
- linear dynamic range
- imaging system noise floor
- fluorescence detection limits
- saturation thresholds
- tissue sensing depth
- fluorescence imaging spatial resolution
- fluorescence lateral spatial resolution
- fluorescence imaging depth of field
- radiometrically standardized performance quantification
Combined with our fluorescence reference targets, the RET provides a fully-traceable, quantitative basis for fluorescence imaging system characterization useful for:
- optimizing image capture parameters
- assessing spectral filtering performance
- predicate device comparisons for equivalency justification (e.g., 510(k) submission)
- establishing Imaging system performance repeatability ahead of production
- system comparisons against a traceable reference standard
- absolute radiometric imaging performance characterization
- managing fluorescence target photobleaching during high-volume use.
All quantitative reference kits include an electronic Certificate of Verification (.PDF). and supporting calibration files (.XLS) for serialized RET targets.
The Quantitative Q800 Reference Set streamlines phantom characterization time while providing reproducible, traceable performance data to support your design verification, validation testing, predicate device equivalency, and regulatory documentation requirements.
Our open-source QUEL-QAL python library helps you generate standardized performance reports from raw image data, enabling consistent benchmarking across development cycles and facilitating objective system-to-system comparisons. Video tutorials for using QUEL-QAL can be found in this playlist on our YouTube page.
To help maintain the performance and longevity of the fluorescent reference targets, we recommend minimizing prolonged light exposure. For absolute radiometric benchmarking of system performance, we recommend using the RET target included in this set.