Q800 Depth Sensitivity Target
9-well fluorophore depth target for fluorescence sensitivity performance characterization in tissue. Includes electronic Certificate of Verification (.PDF).
Fluorophore: Q800-01 (Peak Ex/Em: 775/787 nm)
Suitable for systems optimized for 800 nm channel fluorophores (pafolacianine, OTL38, IRDye® 800CW)
Depth Material: 0.5-6 mm depth
Optical Properties measured at 𝛌 = 800 nm
- Absorption (µₐ) = 0.01 mm⁻¹
- Reduced scattering (µ′ₛ) = 1.5 mm⁻¹
Fluorescent Material: Q800-01 = 1000 nM
Optical Properties measured at 𝛌 = 800 nm
- Absorption (µₐ) = 0.04 mm⁻¹
- Reduced scattering (µ′ₛ) = 1.2 mm⁻¹
Standard Size Dimensions:
- 50 x 50 x 20 mm
- ⦰ 10 mm feature size
Mini Size Dimensions:
- 25 x 25 x 20 mm
- ⦰ 5 mm feature size
Compatible with our open-source QUEL-QAL image analysis library.
Fluorophore: Q800-01 (Peak Ex/Em: 775/787 nm)
Suitable for systems optimized for 800 nm channel fluorophores (pafolacianine, OTL38, IRDye® 800CW)
Optical Properties:
Top: 0.5-6mm tissue-equivalent material non-fluorescent material
- Absorption (µₐ) = 0.01 mm⁻¹ at 800 nm
- Reduced scattering (µ′ₛ) = 1.5 mm⁻¹ at 800 nm
Bottom: 8 wells 600 nM Q800-01
- Absorption (µₐ) = 0.04 mm⁻¹ at 800 nm
- Reduced scattering (µ′ₛ) = 1.2 mm⁻¹ at 800 nm
Negative control has 6mm of top material
Dimensions:
- (Standard): 50x50x20 mm | ⦰10 mm feature size
- (Mini): 25x25x20 mm | ⦰5 mm feature size
Our fluorescence depth sensitivity targets align with the FDA Draft Guidance and AAPM TG311 Report on standardizing benchmarks for fluorescence surgical imaging system performance.
The Q800 Depth Sensitivity Target mimics clinically-relevant imaging conditions with 600-nM Q800 blocks beneath 0.5–6 mm of tissue-mimicking material plus a negative reference well. The depth target helps you:
- characterize imaging depth
- optimize image capture parameters
- assess spectral filtering performance
- compare predicate devices (e.g., 510(k) submission)
We provide a getting started guide, directions for use, and practical tutorials on how to get the most out of your depth sensitivity target for your application.
Our open-source QUEL-QAL python library helps you generate standardized performance reports from raw image data, enabling consistent benchmarking across development cycles and facilitating objective system-to-system comparisons. Video tutorials for using QUEL-QAL can be found in this playlist on our YouTube page.
To help maintain the performance and longevity of the fluorescent reference targets, we recommend minimizing prolonged light exposure. For absolute radiometric benchmarking of system performance, we recommend using the RET target.